題 名 | 以氯化芐二甲烴銨浸泡體溫計效果之評估 |
---|---|
作 者 | 范瑞麟; | 書刊名 | 醫學研究 |
卷 期 | 8:4 1988.01[民77.01] |
頁 次 | 頁249-254 |
分類號 | 415.13 |
關鍵詞 | 二甲烴銨; 效果; 浸泡; 氯化芐; 體溫計; |
語 文 | 中文(Chinese) |
中文摘要 | 本研究為評估病房使用之體溫計是否會因護理常規消毒方法,即浸泡於1:1000苄氯化二甲烴銨(Benzalkonium chloride)消毒液中,而遭受汚染。於總消毒後,第一、三、五及七日採檢體做細菌培養。結果發現如消毒容器内尚置入棉球,體溫計予病人使用後第三日即有25%被細菌汚染,第五天則達70%之汚染率。如未置入棉球於消毒容器,汚染率於擦拭體溫計,則汚染率降至8.6%。 |
英文摘要 | The purpose of this study is to evaluate the rountine distinfectant technique of the thermometer immersed in the 1:1000 Zephiran container and used in the ward. The specimens were collected on the day that disinfection procedure was taken and 1, 3, 5 and 7 days after. The results revealed that the rate of contamination of the thermometer after been used by the patients was 25% on the day 3, and 70% on the day 5, if a cotton ball was previously inserted into the container. The contamination rate dropped to 10% on day 3, and 21% on day 5 if no cotton ball was inserted. Most of bacteria isolated were gram-negative bacilli. Under the same condition, the rate of contamination was decreased to 6.6% if the thermometer was previously wiped with alcohol sponge. |