查詢結果分析
相關文獻
- Full-Thickness Burn Injury Resulted from Magnetic Resonance Imaging with Electrocardiographic Monitoring Leads--A Case Report
- 硬脊膜外麻醉導致下半身癱瘓:病例報告
- Gadolinium-Enhanced MRI in Neuro-Behcet Disease
- Magnetic Resonance Images of Neuronal Migration Anomalies
- 擴散影像在腦缺血及梗塞診斷上的應用價值
- Lumbar Synovial Cysts: Magnetic Resonance Imaging Evaluation
- Orbital Invasion in Nasopharyngeal Carcinoma: Evaluation with Computed Tomography and Magnetic Resonance Imaging
- Primary Cardiac Tumors: Report of Two Cases
- Magnetic Resonance Imaging Features of Neuro-Behcet's Disease: A Case Report
- A Rotational Phantom for Evaluation of MR-Compatible Biopsy Devices: Experiment and Preliminary Observations
頁籤選單縮合
題 名 | Full-Thickness Burn Injury Resulted from Magnetic Resonance Imaging with Electrocardiographic Monitoring Leads--A Case Report=磁振造影檢查時以心電圖監測導致全層皮膚灼傷之個案報告 |
---|---|
作 者 | 劉明偉; 張立言; | 書刊名 | 中華民國整形外科醫學會雜誌 |
卷 期 | 9:2 2000.06[民89.06] |
頁 次 | 頁134-137 |
分類號 | 416.1 |
關鍵詞 | 磁振造影; 心電圖監測; 皮膚灼傷; Burn injury; Magnetic resonance imaging; Electrocardiographic monitoring leads; |
語 文 | 英文(English) |
中文摘要 | 磁振造影是目前臨床診斷之重要工具之一,特別是在檢查中樞神經系統以及軟組織病灶時,磁振造影是不可或缺的診斷利器。然而在操作磁振造影儀器之前,工作人貝必須考量其安全性以避免醫源性之併發症。含鐵而具有磁性的物質在電磁場下會產生電流(eddy current)而放熱的現象,很可能是磁振造影檢查導致灼傷的主要原因。因此,患者身上接觸磁振造影儀器附件或者各種生理監測裝置時,在進行檢查前必須瞭解其是否具有危險性以避免灼傷的發生。由於磁振造影檢查而導致灼傷的個案在台灣未曾被報導過,本文在此提出一患者在磁振造影檢查時因心電圖同步監測而導致胸部全層皮層灼傷之個案報告,提供各位參考並希望能避免此併發症的發生。 |
英文摘要 | Magnetic resonance imaging (MRI) is an important diagnostic tool. MRI is indispensable in diagnosis of the central nervous system and soft tissue lesions. Prior to a clinical MRI examination, various safety considerations should be emphasized to avoid the occurrence of iatrogenic complications. Burn injury resulted from MRI is a rare complication and which has never been reported in Taiwan. Although the cause of the MRI induced burn remains theoretical, the electrical current (eddy current) induced by the magnetic and radiofrequency fields may be the major cause of the injury. Therefore, the MRI accessories or the physiologic monitoring devices should be thorough tested and determined to be safe for patient during MRI procedure. We report a case that a patient under general anesthesia suffered from full-thickness burns due to the use of electrocardiographic (ECG) monitoring leads during an MRI examination. |