題名 | Calibration for the Estimation of Thorium to Uranium Concentration Ratios in Samples by CR-39 and LR-115 Detectors=使用CR-39與LR-115偵檢器進行樣品中釷鈾濃度比例測定之校準研究 |
---|---|
作者姓名(外文) | Moharram,B. M.; | 書刊名 | 核子科學 |
卷期 | 36:5 1999.10[民88.10] |
頁次 | 頁343-348 |
分類號 | 449.62 |
關鍵詞 | CR-39偵檢器; LR-115偵檢器; 釷鈾濃度比例測定; Calibration; Concentration ratio; CR-39; LR-11; Thorium; Uranium; |
語文 | 英文(English) |
英文摘要 | The paper reports on the analysis for the calibration to get an experimental curve which gives the relation between C/C concentration ratio and the ratio of track density on CR-39 to that on LR-115. A group of homogeneous mixtures in different known mixing ratios of monazite and zirconia were prepared. The mixed samples in the form of pellets were irradiated in close contact with small pieces of chemically pre-etched CR-39 detectors on one side and LR-115 on the other side of each test sample. The relation between thorium to uranium concentration ratios C/C versus the ratio of track densities of CR-39 to LR-115 detectors has been obtained as an experimental curve. The obtained experimental curve shows a limited agreement with the calculated curve. A good agreement has been extended until C/C ∼8. Present work was applied on two mixtures of soil and coal fly ash to get the component mixing-ratios in a reasonable accuracy. |