頁籤選單縮合
題名 | Direct Determination of Total Chromium in Sea Water by Graphite Furnace Atomic Absorption Spectrometry with Multiple Injections= |
---|---|
作者 | 謝玲玲; 洪瑛鍈; 藍啟仁; Hsieh, Ling-ling; Hung, Ying-ying; Lan, Chi-ren; |
期刊 | Journal of the Chinese Chemical Society |
出版日期 | 19930600 |
卷期 | 40:3 1993.06[民82.06] |
頁次 | 頁241-244 |
分類號 | 340 |
語文 | eng |
關鍵詞 | Total chromium; Sea water; Multiple injections; Graphite furnace atomic absorption spectrometry; |
英文摘要 | The method of multiple injections coupled with graphite furnace and atomic absorption spectrometry is applied to direct determination of total chromium at extremely small concentrations in samples of open sea water. The method involves the in situ preconcentration of trace chromium in sea water onto a pyrocoated graphite tube by multiple injections (up to 4 x 90 μ L ) prior to analysis. The selected ashing temperature is up to 1800 ℃, and the background signal caused by high sea-salt contents will be markedly reduced when the ashing time is prolonged up to 200s. A CASS-2 reference sea water has been used as a quality control sample in the analyses of samples of open sea water samples; the results were found to agree with the certified value. Due to the use of a relatively large volume (up to 0.36 mL) of sample for direct determination of trace chromium in sea water, the detection limit for chromium is 0.057 ppb. |
本系統之摘要資訊系依該期刊論文摘要之資訊為主。